Specifications & operating manual
English · 31 pages · PDF · 1.3 MB
Download manual ↓ADSP-HV80+ — a multi-channel high-voltage power supply with filament drive for selected HITACHI CD-SEM systems.

Source: manual V2609E07, with manufacturer-confirmed corrections.
| AC input | 100–230 V AC · 1 phase · 50/60 Hz · 40 W |
|---|---|
| HV output accuracy | < ±0.02% of setting, after warm-up |
| HV output ripple | < 40 mV (p-p) |
| Filament drive | 0–2.85 A DC · constant current |
| Dimensions (D × W × H) | 435 × 230 × 165 mm |
| Weight | Approx. 6.5 kg unit / approx. 8.3 kg including HV cable and connector |
| Temperature coefficient | < 0.01% / °C |
|---|---|
| Filament current accuracy | < ±0.2% of setting, after warm-up |
| Communication | RS-232 · DB25F · optically isolated |
| Protection | Overload, short circuit, vacuum interlock; If > 3 A; Ie > 511 μA |
| HV cable | 2.2 m · shielded · five HV wires plus separate ground wire |
| Cable construction | PVC outer sheath around the harness; silicone outer jacket on each of the five individual HV wires |
| Operating conditions | Indoor use · 15–30 °C · 35–50% RH, non-condensing |
| Enclosure protection | IP20 |
| Cooling | Forced air; minimum 50 mm clearance on both sides and above recommended |
Built-in V1/V2 characteristic curves support the following equipment models. Tell us your tool model and existing HV unit part number so we can confirm the required configuration.
* CG4000 requires a separate software upgrade package.
Optically isolated RS-232 (DB25F), vacuum interlock LINK and a dedicated high-voltage connector. The AC input connection depends on the tool series; an adapter cable is provided for S-92XX and S-93XX.
Replacement applications listed in the manual
567-5101 · 560-6590 · 569-6592
English · 31 pages · PDF · 1.3 MB
Download manual ↓Applications and demonstrations on our YouTube channel. Links open YouTube in a new tab; no YouTube content is loaded on this page.
View application videos ↗For professional use. Installation by qualified personnel only. Read the manual and observe the specified operating conditions.